We are counted among the renowned supplier of the Fresenius Hemodialysis Machine. These dialysis machines are designed and developed utilizing quality-approved metals in compliance with industry standards of quality.

Features:

  • Easy operation
  • Energy efficient
  • Smooth finish
  • Longer functional life
Specifications:
  • Dimensions: 1330 x 495 x 340 mm
  • Weight: approx. 80 kg

Water Supply:

  • Inlet Pressure: 1.5-6.0 bar
  • Water Inlet Temperature: 5-30oC

Electrical Data:

  • Power Supply: 230 v - 10 % A, 50 Hz
  • Current Consumption: max 9 A

Extra Corporeal Circuit:

  • Arterial Pressure: Display Range: -300 mmhg to +280 mmhg
  • Accuracy: +/- 10 mmhg
  • Resolution: 20 mmhg

Trans Membrane Pressure:

  • Display range: -60 mmhg to + 500 mmhg
  • Resolution: 20 mmhg

Arterial Blood Pump:

  • Blood Flow Range: 10-600 ml/min
  • Accuracy: +/- 10%
  • Air Bubble Detector: By ultra sound transmission,
  • Additional optical monitoring

Heparin Pump:

  • Delivery Range: 0 to 10 ml
  • Bolus Function: max 5 ml per bolus
  • Syringe Size: 20 ml, 30 ml and 50 ml

Dialysate Fluid Circuit:

  • Select able: 0 300 - 500 ml/min
  • Dialysate Fluid Temperature: Selectable: 35-39oC
Dialysis Fluid Conductivity:
  • Range: 12.8 to 15.7 ms/cm. accuracy : +/- 0.1 ms/cm

Sodium Concentration Dialysis Fluid:

  • Default Mixing Ratio: 1 + 34
  • Range: - 13 to +13 mmol/bicarbonate

Bicarbonate Concentration:

  • Default Mixing Ratio: 1 + 27.6
  • Range: - 8 to + 8 mmol/l bicarbonate
  • Balancing Accuracy: +/- 0.1 % of the dialysate flow

Ultra Filtration:

  • UF Rate: 0 to 4 lph
  • Accuracy: =/- 0.5%
  • Allowed Dialyses UF Factor: unlimited
  • Parameter Displayed: uf goal/uf time/uf rate/uf volume
  • Blood Leak Detector: Sensitivity:

Disinfection and Cleaning Programs:

  • Rinse: Temperature: 37oC/600 ml/min
  • Hot Rinse: Temperature: 85oC/450 ml/min
  • Cleaning Portal: Temperature: 37
  • Hot Disinfection: Temperature: 85oC/450 ml/min
  • Sodium Profiling: Yes
  • Ultra Filtration Profiling: Yes